Finger-prick and upper-arm sampling: validated against venous blood.
When we measure blood values from a few drops of blood, we want to know whether the result matches blood drawn from a vein. So each person gave two samples, and we measured both in our laboratory in Eugendorf, Austria. This page lists all three comparisons we have run so far: two with blood from the fingertip and one with blood from the upper arm.
How we compare
- Same person, two samples. Each person gives blood from a vein and blood from the fingertip or the upper arm. We measure both samples on the same analyser.
- One limit per blood value. For every value we set in advance how far apart the two results may be. These limits come from published tables of allowable measurement error. A pair within the limit counts as matching.
- We count. For every blood value the tables show how many pairs were within the limit. Values that did not pass are listed too.
Comparison 1: fingertip against vein, 6 blood values (2023)
40 people gave blood from a vein and blood from the fingertip at the same time. We measured active vitamin B12, magnesium, vitamin D, LDL and HDL cholesterol and triglycerides.
For magnesium, vitamin D, LDL and HDL every pair was within the limit. For active vitamin B12 it was 33 of 37, for triglycerides 35 of 40. One person differed strongly in B12 and triglycerides; their finger value for B12 was almost twice the vein value.
| Blood value | Pairs within the limit | Share | Allowed difference |
|---|---|---|---|
| Active vitamin B12 | 33 of 37 | 89% | 30% |
| Magnesium | 40 of 40 | 100% | 25% |
| Vitamin D | 40 of 40 | 100% | 15% |
| LDL cholesterol | 40 of 40 | 100% | 30% |
| HDL cholesterol | 40 of 40 | 100% | 30% |
| Triglycerides | 35 of 40 | 88% | 25% |
Comparison 2: fingertip against vein, hormones and CRP (2023/24)
This comparison covered hormones, the thyroid and CRP. Blood from the fingertip went into a small tube and was measured on day five, blood from the vein on day one. Some people gave samples more than once, so here we count sample pairs, not people.
For DHEA-S, TSH, free T3, oestradiol and testosterone in men every pair was within the limit. For free T4, FSH, LH, SHBG, progesterone and CRP one or two pairs were outside. Two values did not pass: cortisol and testosterone in women.
| Blood value | Pairs within the limit | Share | Allowed difference |
|---|---|---|---|
| DHEA-S | 63 of 63 | 100% | 22% |
| TSH | 70 of 70 | 100% | 30% |
| Free T3 | 71 of 71 | 100% | 20% |
| Free T4 | 60 of 62 | 97% | 15% |
| FSH | 63 of 65 | 97% | 25% or 4 IU/L |
| LH | 60 of 62 | 97% | 25% or 2 IU/L |
| SHBG | 67 of 68 | 99% | 21% |
| CRP | 65 of 66 | 98% | 68% |
| Oestradiol (women only) | 35 of 35 | 100% | 30% |
| Progesterone (women only) | 33 of 34 | 97% | 25 to 30% |
| Testosterone, men | 30 of 30 | 100% | 25 to 30% |
| Testosterone, women | 22 of 28 | 79%, did not pass | 0.3 ng/mL |
| Cortisol | 59 of 66 | 89%, did not pass | 25 to 30% |
- Cortisol did not pass. One possible reason: cortisol rises sharply in the morning and then falls again. If the finger sample took longer, the value may already have dropped in the meantime.
- Testosterone in women did not pass: 6 of 28 pairs were outside the limit of 0.3 ng/mL, the largest difference was 0.5 ng/mL. From the fingertip, testosterone therefore counts as tested for men only.
- Oestradiol and progesterone apply to women only. In men there were too many differences and too few samples.
Comparison 3: upper arm against vein, 17 blood values (2026)
23 healthy adults, 11 men and 12 women aged 22 to 60, gave blood from a vein and blood from the upper arm on three mornings. On the upper arm we collected the blood with a small device that is placed on the skin. We processed both samples the same way and measured them in the same run.
For 14 of the 17 values every pair was within the limit. For TSH it was 21 of 23, for LH 22 of 23, for triglycerides 21 of 23. The reasons are below the table. On average, the upper-arm values for most blood values were less than 3 percent away from the vein values.
| Blood value | Pairs within the limit | Allowed difference | Average difference |
|---|---|---|---|
| LDL cholesterol | 23 of 23 | 20% | −1.3% |
| HDL cholesterol | 23 of 23 | 20% or 6 mg/dL | −1.0% |
| Triglycerides | 21 of 23 | 15% or 17.7 mg/dL | −4.2%* |
| CRP | 23 of 23 | 30% or 0.1 mg/dL | −2.8% |
| Magnesium | 23 of 23 | 15% | −2.2% |
| Active vitamin B12 | 23 of 23 | 25% | −1.5% |
| Vitamin D | 23 of 23 | 15% or 3.6 ng/mL | −0.7% |
| TSH | 21 of 23 | 20% or 0.2 mIU/L | −8.9% |
| Free T3 | 23 of 23 | 20% or 0.46 pg/mL | +0.3% |
| Free T4 | 23 of 23 | 15% or 3 pg/mL | +2.6% |
| FSH | 23 of 23 | 18% or 2 IU/L | +2.0% |
| LH | 22 of 23 | 20% or 1.5 IU/L | −7.2% |
| Oestradiol | 23 of 23 | 30% | +7.4% |
| Progesterone | 23 of 23 | 25% | +1.8% |
| SHBG | 23 of 23 | 12% | −0.2% |
| DHEA-S | 23 of 23 | 12% or 44 µg/dL | −5.5% |
| Testosterone | 23 of 23 | 30% or 0.2 ng/mL | −7.8% |
* Triglycerides without the two difficult draws (21 pairs). Where it says “or”, the larger of the two applies.
- Triglycerides: Two pairs were outside. For both people the upper-arm draw was difficult. One sample was very fatty, in the other the blood came 106 to 141 minutes after the vein draw. Without these two, all 21 pairs were within the limit. For the planned kit, triglycerides are therefore to be measured from fingertip blood when the upper-arm draw is difficult.
- TSH: One pair was just outside (21 percent difference where 20 was allowed), one came from the difficult draw with more than 100 minutes in between. TSH falls by itself over the morning, and the upper-arm blood was always taken after the vein blood.
- LH: One pair was outside. The body releases LH in bursts, so the value can change by itself between the two draws.
What the comparisons do not show
- The upper arm was always sampled after the vein, on average 22 minutes later. Part of the differences comes from this gap and not from where the blood was taken.
- With 21 to 71 pairs per value, the comparisons are on the small side. Professional guidance for such comparisons asks for at least 20 samples.
- The results apply to our tubes, our procedures and our analyser. With other equipment they may turn out differently.
- The comparisons test whether two ways of taking blood give the same result. They do not show what a single value means for a person’s health.
- We completed the upper-arm comparison in November 2026. The internal release for routine laboratory use is still in progress.
Frequently asked questions
Does blood from the fingertip give the same values as blood from a vein?
How did the comparison with upper-arm blood work?
What does “within the limit” mean?
Why is the upper arm sampled later than the vein?
Does this page tell me whether my blood values are fine?
Read more
This page summarises comparison measurements from our own laboratory and is for background information. It is not medical advice, diagnosis or a treatment recommendation. Our products are available through our partner companies. This page was created with the help of AI. Who writes here and where the sources come from.